FieldMaxII-TOP User Manual
50
The trigger is synchronous with the leading edge of the
pulse, but the actual peak is determined algorithmically
by sampling the input signal near the trigger. From the
trigger point forward, the algorithm searches for peaks
and from the trigger point back, it searches for a baseline.
Hz Display
Mode
If desired, pulse frequencies may be displayed when a
pyroelectric sensor is attached to the meter. For more
information, refer to “Hz” on page 38.
Manual
Ranging Mode
(thermopile or optical sensors only) Manual ranging
requires the user to select the range. For more informa-
tion about manual ranging, refer to “Up and Down
Arrows” on page 40.
Thermopile-
and Optical-
Specific
Information
Information in this section pertains exclusively to ther-
mopile and optical sensors. If you are using a pyroelec-
tric sensor, refer to “Pyroelectric-Specific Information”
on page 48.
Auto vs. Manual
Ranging Mode
Auto Ranging (Auto) enables FieldMaxII-TOP to auto-
matically select the range (gain) when a thermopile or
optical sensor is attached to the meter. While in Auto
Ranging mode, the tuning needles represent a zero-to-
full scale movement (as compared to Tuning mode,
where the zones overlap). Refer to Figure 16.
Przeglądanie stron 63