Coherent FieldMaxII-TOP Instrukcja Użytkownika Strona 48

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 118
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 47
FieldMaxII-TOP User Manual
34
display the wavelength value. The available wavelength
range is 1.00 to 99,999 nm. The actual range is sensor-
dependent.
If Wavelength compensation information is not
programmed into the sensor, you will not be able to
change the wavelength data value.
Area
This mode allows the measurement of laser energy in
terms of fluence, and laser power in terms of average
power density.
Area calculation returns an average power density
reading for both Flat and Gaussian profiles.
The parameter is entered as a diameter and assumes a
circular beam or aperture. The range for Area mode is
0.01 to 999.99 mm.
Pressing the Area button will:
Toggle the state of Area Correction mode, if the
button is pressed for less than one second. Refer to
“Area Correction and Zeroing” on page 35 for
more information about Area Correction mode.
Enter Edit mode and select the beam diameter, if
Setup/Local is pressed beforehand.
View the beam diameter, if the button is pressed for
one second or more.
Przeglądanie stron 47
1 2 ... 43 44 45 46 47 48 49 50 51 52 53 ... 117 118

Komentarze do niniejszej Instrukcji

Brak uwag