Operation
39
until the procedure ends. The zero procedure is immedi-
ately terminated if the sensor is disconnected or if an
error is encountered.
Normally, the Zero button is pressed with the laser
blocked from the connected sensor. If a finite power
level is present at the sensor, the instrument will attempt
to null it out. A bad zero fault code appears if a given
power input is too large to null on the sensitive ranges.
If the bad zero fault code appears:
Press any soft button to dismiss the error
Select a new range
Press the Zero button
The secondary function of the Zero button is to manually
start a batch while in Statistics mode.
Power Switch
and Backlight
Toggle Button
The combination Power Switch and Backlight
Toggle button serves the dual purpose of
turning power on/off to the meter, and
toggling the backlight.
When the meter is off, the power-on state is acti-
vated by pressing the button for one second.
Pressing the button for one second while in the
power-on state turns the meter off.
When the meter is in the power-on state, the back-
light state is toggled by pressing the button for less
than one second.
The backlight is always off when power is first
applied to the meter.