FieldMaxII-TOP User Manual
36
Avg
Average mode enables display smoothing, which
suppresses variations in the display reading that can
make it difficult to read.
Display values are averaged either by time (for ther-
mopile or optical sensors), or by pulses (for pyroelec-
tric sensors).
You can change the size of the display smoothing
window. With an attached thermopile or optical sensor,
the window size is 1 to 60 seconds. Using a pyroelectric
sensor, the window size is 2 to 1,000 pulses.
The Avg button is used to:
Toggle the state of Average mode (if the button is
pressed for less than one second).
Enter Edit mode and select the display smoothing
window size (if Setup/Local is pressed before-
hand).
View the display smoothing window size (if the
button is pressed for one second or more).
Atten
Attenuation mode allows you to get true measurements
using an attenuator that has a known attenuation factor.
When Attenuation Correction mode is enabled, the
measured value is adjusted to indicate the measurement
at the attenuator and not the sensor. The range for this
mode is 0.01 to 999.99.