Quick Start
9
Measuring
Energy With a
Thermopile
Sensor
The following figure explains how to take an energy
measurement using a thermopile sensor.
Figure 4. Measuring Energy With a Thermopile Sensor
Press the J/W button to select Watts mode and
then find the appropriate watts range by taking
a couple of sample measurements. The
appropriate range is the lowest range
available that does not give an
overrange error during the test
measurement.
!
Connect a thermopile sensor to
the 25-pin connector.
"
#
Block the beam and then press the
Zero button to set the baseline for
your new measurement.
W (indicates Watts mode)
J
(indicates Joules mode)
Expose the sensor to a laser pulse, take the measurement, and observe the result on the LCD.
%
Press the J/W button to select Joules
mode. The Range (Up and Down
arrows), Zero button, and Auto button
should not be used from this point on.
If an overrange error occurs, the range must be adjusted by returning to the Watts mode.
$
When FieldMax II-TOP is prepared for the first measurement, the TRIG? annunciator displays at the top of
the screen. The TRIG annunciator displays during the measurement, showing that the meter is currently
integrating the reading. The Zero button should not be pressed between measurements.