Operation
41
Selecting Auto mode restarts the batch count used to take
a reading. If the instrument is not in Auto mode, the batch
count must be manually restarted by pressing the Zero
button.
Pressing and holding the Stat button displays the
current instrument mode: Auto or Manual.
For more information on selecting parameters while in
Statistics mode, refer to “Left and Right Arrows” on
page 40.
Invalid Data
A statistical batch requires valid data for every data point
in the batch. If a batch collection of data is in process and
invalid data is measured, the batch is considered contam-
inated and the batch immediately ends with no batch
result computed. If the Restart mode is Auto, a new batch
is immediately restarted. The error that caused the
contaminated batch is displayed.
Energy and
Power
Measure-
ments
Energy (joules) measurements are taken using pyroelec-
tric sensors (for pulsed laser) or thermopile sensors (for
long-pulsed laser). The user interface behaves in a