54
Superior Reliability & Performance • www.Coherent.com • LMC.sales@coherent.com
* Requires 15 VDC power input
83 mm
(0.59 in.)
83 mm
(0.59 in.)
Ø36 mm
(1.42 in.)
22 mm
(0.87 in.)
89 mm
(3.50 in.)
89 mm
(3.50 in.)
Ø50 mm
(1.97 in.)
32 mm
(1.26 in.)
Laser
Cross-
Reference
Index
POWER
& ENERGY
Power
Sensors
Energy
Sensors
Custom
& OEM
BEAM
DIAGNOSTICS
CALIBRATION
& SERVICE
INDICES
Product
Name
Index
Power
& Energy
Meters
25.4 mm
(1.0 in.)
83.8 mm
(3.3 in.)
83.8 mm
(3.3 in.)
35.6 mm
(1.4 in.)
BeamFinder: Use with
LabMax, Ultima, or LaserPAD
(see pages 10-16)
300 mW to 1 kW
OEM Thermopiles
Features
• 300 mW to 1 kW
• Spectrally at from 0.19 μm to 11 μm
• Compact designs
• Air-cooled or water-cooled
• Active area diameters from 19 mm to 50 mm
Model PM150-50B, PM150-50XB, PM1K-36B, BeamFinder
Model PM150-50B PM150-50XB PM1K-36B BeamFinder
1
Wavelength Range (µm) 0.19 to 11 0.15 to 1 0.19 to 11 0.3 to 10.6
Resolution (mW) 30 30 1000 1000
Min. Water ow (gpm) 0.2 0.2 1.0 1.0
Max. Avg. Power (water-cooled) (W) 150 1000
Max. Avg. Power (air-cooled, 5 min.) (W) 80 40 –
Responsivity (typ.) 0.4 mV/W 0.1 mV/W –
Max. Power Density 26 kW/cm
2
2.5 kW/cm
2
Max. Energy Density 0.6 J/cm
2
, 0.5 J/cm
2
,
1064 nm, 10 ns 1064 nm, 10 ns
Response Time (sec.) 5 10
Detector Coating Broadband UV Broadband H
Active Area Diameter (mm) 50 36 35
Dimensions (mm) 89 x 89 x 32 83 x 83 x 22 84 x 84 x 25.4
(3.5 x 3.5 x 1.2 in.) (3.2 x 3.2 x 0.8 in.) (3.3 x 3.3 x 1.0 in.)
Calibration Uncertainty (%) ±1 ±5
Calibration Wavelength (nm) 514 1064 10,600
Cooling Method Water-cooled
Connector Type BNC-terminated LM DB-25
Cable Length (m) – – – 6
Part Number (RoHS) 1098415 1098441 1098333 1098427
Device
Specications
PM150-50B/PM150-50XB
PM1K-36B
BeamFinder
1
BeamFinder incorporates a quadrant thermopile disk that enables the position of the beam to be sensed.
Komentarze do niniejszej Instrukcji