Coherent FieldMaxII-TOP Dokumentacja Strona 34

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 92
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 33
32
Superior Reliability & Performance • www.Coherent.com • LMC.sales@coherent.com
Ø50 mm
(1.97 in.)
230 mm
(9.06 in.)
59 mm
(2.32 in.)
120 mm
(4.73 in.)
Laser
Cross-
Reference
Index
POWER
& ENERGY
Power
Sensors
Energy
Sensors
Custom
& OEM
BEAM
DIAGNOSTICS
CALIBRATION
& SERVICE
INDICES
Product
Name
Index
Power
& Energy
Meters
Features
Water-cooled
Spectrally at from 0.19 μm to 11 μm
1W resolution
50 mm apertures
Model PM1K
These water-cooled sensors are used to measure lasers over 300W average
power output. They are excellent choices for measuring CO
2
and Nd:YAG
lasers. Larger-area versions are available on the next page.
Tap or distilled cooling water is recommended with these sensors – DI
water can not be used. Flow rates are power dependent and range from
0.5 to 4 gallons per minute; pressure depends upon ow rate and ranges
from 3 to 40 PSI (visit product pages at www.Coherent.com/LMC for more
technical details).
Model PM1K PM3K PM5K
Wavelength Range (µm) 0.19 to 11
Power Range (W) 100 to 1000 100 to 3000 100 to 5000
Max. Intermittent Power (<5 min.)(W)
1
3000 5000 10000
Resolution (W) 1
Max. Power Density
2
1 to 10 kW/cm
2
Max. Energy Density 0.6 J/cm
2
, 1064 nm, 10 ns
Response Time (sec.) 30
Detector Coating Broadband
Active Area Diameter (mm) 50
Dimensions (mm) Ø120 x 59 (4.7 x 2.3 in.)
Calibration Uncertainty (%) ±5
Calibration Wavelength (nm) 1064
Cooling Method Water-cooled
Cable Type PM DB-25
Cable Length (m) 2
Part Number 1098392 1098462 1098454
Device
Specications
1
Intermittent power levels may be sustainable for longer than 5 minutes when used with lasers with large diameter, non-Gaussian beam proles.
Monitor closely for coating damage if used longer than ve minutes at higher powers.
2
The damage resistance of the coating is dependent upon the beam size and prole, the average power level, and the water ow rate.
Contact Coherent or your local representative for details related to your application.
PM1K/PM3K/PM5K
100W to 5 kW
Water-Cooled Thermopile Sensors
Przeglądanie stron 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 91 92

Komentarze do niniejszej Instrukcji