Host Interface
67
propget HoldMode([out, retval] VARIANT_BOOL* );
The read only property indicating hold mode state. This
parameter has effect only with a TO meter.
propget LastData([out, retval] double* );
The read only property indicating the last measurement
value. This property is read in response to a “Measure-
mentData” event.
propget LastFault([out, retval] short* );
The read only property indicating the fault code of the
last fault event. This property is read in response to a
“Fault” event. The last fault code may be cleared by
calling the DismissFault() method.
propget LastPeriod([out, retval] double* );
The read only property indicating the last period value.
This property is meaningful only when a pyroelectric
probe is attached. Frequency in Hz is calculated as
125000/LastPeriod.
propget MaxRange([out, retval] double* );
The read only property indicating the maximum full
scale range that the meter will accommodate. This prop-
erty automatically varies as different probes are attached.
propget MaxWavelength([out, retval] long* );
The read only property indicating the maximum wave-
length that the meter will accommodate. This property
automatically varies as different probes are attached.
propget MeasurementMode([out, retval] short* );
The read only property indicating the measurement
mode (Watts or Joules).